# Beer–Lambert law calculator: concentration from absorbance (A = εlc)

> Solve A = εlc for concentration, absorbance, ε or path length, in M, µM or mg/mL, convert %T to absorbance, and pick ε values with sources. Free.

URL: https://www.labcanvas.com/tools/beer-lambert-law-calculator

The Beer–Lambert law says that absorbance is proportional to concentration: A = ε·l·c, with ε the molar absorption coefficient (M⁻¹ cm⁻¹), l the path length (cm) and c the concentration (mol/L). So c = A/(ε·l): an NADH solution with A340 = 0.415 in a 1 cm cuvette (ε = 6220 M⁻¹ cm⁻¹) is 66.7 µM. Choose which of the four to calculate, enter the other three in any units (M, mM, µM, mg/mL, µg/mL; cm or mm; ε per mole or per gram, joined by the molar mass), or enter the transmittance instead of the absorbance (A = 2 − log₁₀ %T).

## Quick answer

- **What it does:** Solves the Beer–Lambert law A = ε·l·c for the concentration, the absorbance, the absorption coefficient or the path length, with molar or mass units and %T.
- **Formula:** A = ε·l·c;  c = A/(ε·l);  A = −log₁₀ T = 2 − log₁₀(%T);  c (g/L) = c (mol/L) × M
- **Example:** NADH at 340 nm, ε = 6220 M⁻¹ cm⁻¹, A = 0.415 in a 1 cm cuvette: c = 0.415 / (6220 × 1) = 6.672 × 10⁻⁵ M = 66.72 µM. A transmittance of 50 % is A = 0.301.

Free, in the browser, no account needed. Updated 2026-11-23.

## Features

- Solve A = εlc for any one of concentration, absorbance, ε or path length
- Concentrations in M, mM, µM, nM, g/L, mg/mL, µg/mL, ng/µL, mg/L or µg/L; path length in cm, mm or µm
- ε as M⁻¹ cm⁻¹, mM⁻¹ cm⁻¹, per mass (L g⁻¹ cm⁻¹, mL µg⁻¹ cm⁻¹) or E1%, with the molar mass to convert between molar and mass units
- Absorbance or %transmittance in (A = 2 − log₁₀ %T), and both out
- Molar absorption coefficients of NADH, TNB (Ellman’s), 4-nitrophenolate and the protein chromophores, each with its primary source
- Warnings outside the comfortable absorbance range (below 0.1 or above 1)

## How to use it

1. **Choose what to calculate**: Usually the concentration from a measured absorbance; you can also get the expected absorbance, ε from a standard, or the path length.
2. **Enter what you know**: The absorbance (or %T), ε (pick one from the list or type it, with its unit), the path length (1 cm for a standard cuvette) and, if needed, the concentration, with units.
3. **Add the molar mass if needed**: When ε is per mole and you want mg/mL (or the reverse), enter the molar mass to convert.
4. **Read the answer**: Read the result with the concentration in molar and mass units and the transmittance, and copy the value or the calculation.

## At a glance

| | |
| --- | --- |
| Formula | \(A = \varepsilon l c\); \(A = 2 - \log_{10}(\%T)\) |
| Solves for | Concentration, absorbance, ε or path length |
| Units | M to nM, g/L to µg/L (molar mass joins them); cm, mm, µm; ε per mole, per mass or E1% |
| ε values | NADH, TNB, 4-nitrophenolate, Trp/Tyr/cystine, each with its primary source |
| Cautions | Absorbance below 0.1 or above 1 flagged; one absorbing species, monochromatic light |

## How it works

$$A = \log_{10}\frac{I_0}{I} = \varepsilon\, l\, c$$

**Absorbance.** The (decadic) absorbance is \(A = \log_{10}(I_0/I) = -\log_{10} T\), with \(T = I/I_0\) the transmittance (IUPAC Gold Book: absorbance). In percent, \(A = 2 - \log_{10}(\%T)\): 10 % transmitted is A = 1, 1 % is A = 2.

**The law.** For a dilute solution of one absorbing species, absorbance is proportional to the path length and the amount concentration, \(A = \varepsilon l c\) (IUPAC Gold Book: Beer–Lambert law). The molar (decadic) absorption coefficient \(\varepsilon\) is \(A/(lc)\), in L mol⁻¹ cm⁻¹ = M⁻¹ cm⁻¹ (IUPAC Gold Book: molar absorption coefficient); it depends on the wavelength, the solvent, the pH and the temperature, so it is quoted with them. With a mass concentration (g/L) the same law uses a specific (mass) absorption coefficient in L g⁻¹ cm⁻¹ = mL mg⁻¹ cm⁻¹; the two are related by the molar mass \(M\): \(\varepsilon_\text{mass} = \varepsilon / M\). E1% is the absorbance of a 1 % (10 mg/mL) solution in 1 cm, 10 × the mass coefficient. The calculator converts every input to mol/L or g/L, cm and M⁻¹ cm⁻¹ or L g⁻¹ cm⁻¹, solves, and converts back.

**Where it fails.** The law assumes monochromatic light, a dilute solution and a single species that does not associate, dissociate or react as the concentration changes; deviations come from chemical equilibria and from polychromatic light (Swinehart 1962), and at high absorbance from stray light. In practice readings are most precise at moderate absorbances, roughly 0.1 to 1: below that the signal is close to the blank’s noise, and above it so little light reaches the detector (under 10 %) that stray light flattens the response. Dilute concentrated samples or use a shorter path, and check linearity with a standard curve (for example with the [protein assay standard curve](https://www.labcanvas.com/tools/protein-assay-standard-curve) tool).

**Nucleic acids and proteins.** For DNA and RNA from A260 (a dsDNA solution of about 50 µg/mL has A260 = 1 in 1 cm) use the [DNA concentration calculator](https://www.labcanvas.com/tools/dna-concentration-calculator), which also gives purity ratios and copy numbers. For a protein, ε at 280 nm can be predicted from its tryptophan, tyrosine and cystine content (Pace et al. 1995); the [protein parameter calculator](https://www.labcanvas.com/tools/protein-parameter-calculator) does it from the sequence.

## Worked examples: NADH, a thiol assay, a protein and %T

- **NADH from A340.** A dehydrogenase reaction mix reads A340 = 0.415 in a 1 cm cuvette, against a blank without NADH. With \(\varepsilon_{340} = 6220\ \mathrm{M^{-1}\,cm^{-1}}\): \(c\) \({}= 0.415 / (6220 \times 1)\) \({}= 6.672 \times 10^{-5}\ \mathrm{M}\) \({}= \mathbf{66.72\ \upmu M}\). The same in a microvolume reader with a 1 mm path would read A = 0.0415, too close to the blank to be precise.
- **Thiols by Ellman’s assay.** After DTNB reacts with a sample, A412 = 0.283 above the reagent blank in 1 cm. Each thiol releases one TNB²⁻ (\(\varepsilon_{412} = 14{,}150\ \mathrm{M^{-1}\,cm^{-1}}\)), so the thiol concentration in the cuvette is \(0.283/14{,}150 = \mathbf{20.0\ \upmu M}\); multiply by the dilution of the sample into the assay.
- **Expected absorbance of a phosphatase product.** 25 µM 4-nitrophenol in 10 mmol/L NaOH (\(\varepsilon_{401} = 18{,}380\ \mathrm{M^{-1}\,cm^{-1}}\)) gives \(A = 18{,}380 \times 1 \times 25 \times 10^{-6} = \mathbf{0.4595}\).
- **A protein in mg/mL from a molar ε.** A protein with \(\varepsilon_{280} = 43{,}824\ \mathrm{M^{-1}\,cm^{-1}}\) and \(M = 66{,}430\ \mathrm{g/mol}\) has a mass coefficient of \(43{,}824/66{,}430\) \({}= 0.6597\ \mathrm{mL\,mg^{-1}\,cm^{-1}}\) (E1% = 6.60): a 1 mg/mL solution reads A280 = 0.660, and A280 = 0.33 means 0.50 mg/mL.
- **Transmittance.** An old instrument reads 50 %T: \(A = 2 - \log_{10} 50 = \mathbf{0.301}\). A = 1.5 lets only \(10^{-1.5} = 3.2\ \%\) of the light through.


## Molar absorption coefficients with their sources

Values to pick in the calculator. Each depends on the conditions given; use your own calibration when the buffer, pH or temperature differ.

| Species | λ (nm) | ε (M⁻¹ cm⁻¹) | Notes | Source |
| --- | --- | --- | --- | --- |
| NADH (and NADPH) | 340 | 6,220 | Reduced band of the nicotinamide coenzymes; the oxidised forms do not absorb at 340 nm. | Horecker & Kornberg 1948, doi:10.1016/S0021-9258(18)57268-9 |
| TNB²⁻ (Ellman’s assay for thiols) | 412 | 14,150 | The coloured anion released when DTNB reacts with a thiol; one TNB per SH. | Riddles, Blakeley & Zerner 1979, doi:10.1016/0003-2697(79)90792-9 |
| 4-Nitrophenol in 10 mmol/L NaOH (4-nitrophenolate) | 401 | 18,380 | Measured at 25 °C; the product of phosphatase and glycosidase assays with 4-nitrophenyl substrates. | Bowers et al. 1980, doi:10.1093/clinchem/26.6.724 |
| Tryptophan residue in a folded protein | 280 | 5,500 | Average for proteins in water; ε(280) = 5500·nTrp + 1490·nTyr + 125·nCystine. | Pace et al. 1995, doi:10.1002/pro.5560041120 |
| Tyrosine residue in a folded protein | 280 | 1,490 | As above. | Pace et al. 1995, doi:10.1002/pro.5560041120 |
| Cystine (disulfide bond) in a folded protein | 280 | 125 | As above. | Pace et al. 1995, doi:10.1002/pro.5560041120 |

For double-stranded DNA, A260 = 1 corresponds to about 50 µg/mL (a mass coefficient of 0.020 mL µg⁻¹ cm⁻¹): see the [DNA concentration calculator](https://www.labcanvas.com/tools/dna-concentration-calculator) for DNA, RNA and oligos.

## References

- IUPAC. Absorbance. In: *Compendium of Chemical Terminology* (the Gold Book). [doi:10.1351/goldbook.A00028](https://doi.org/10.1351/goldbook.A00028)
- IUPAC. Molar absorption coefficient. In: *Compendium of Chemical Terminology* (the Gold Book). [doi:10.1351/goldbook.M03972](https://doi.org/10.1351/goldbook.M03972)
- IUPAC. Beer–Lambert law. In: *Compendium of Chemical Terminology* (the Gold Book). [doi:10.1351/goldbook.B00626](https://doi.org/10.1351/goldbook.B00626)
- Swinehart DF. The Beer–Lambert law. *J Chem Educ* 1962;39:333. [doi:10.1021/ed039p333](https://doi.org/10.1021/ed039p333)
- Horecker BL, Kornberg A. The extinction coefficients of the reduced band of pyridine nucleotides. *J Biol Chem* 1948;175:385–390. [doi:10.1016/S0021-9258(18)57268-9](https://doi.org/10.1016/S0021-9258(18)57268-9)
- Riddles PW, Blakeley RL, Zerner B. Ellman’s reagent: 5,5′-dithiobis(2-nitrobenzoic acid), a reexamination. *Anal Biochem* 1979;94:75–81. [doi:10.1016/0003-2697(79)90792-9](https://doi.org/10.1016/0003-2697(79)90792-9)
- Bowers GN, McComb RB, Christensen RG, Schaffer R. High-purity 4-nitrophenol: purification, characterization, and specifications for use as a spectrophotometric reference material. *Clin Chem* 1980;26:724–729. [doi:10.1093/clinchem/26.6.724](https://doi.org/10.1093/clinchem/26.6.724)
- Pace CN, Vajdos F, Fee L, Grimsley G, Gray T. How to measure and predict the molar absorption coefficient of a protein. *Protein Sci* 1995;4:2411–2423. [doi:10.1002/pro.5560041120](https://doi.org/10.1002/pro.5560041120)


## Related calculators

- [Concentration converter](https://www.labcanvas.com/tools/concentration-converter): Convert between M, mg/mL, % w/v, % w/w, ppm, molality and normality.
- [DNA concentration & copy number calculator](https://www.labcanvas.com/tools/dna-concentration-calculator): DNA, RNA and oligo concentration from A260, purity ratios, and ng/µL to nM and copies per µL.
- [Protein parameter calculator](https://www.labcanvas.com/tools/protein-parameter-calculator): Molecular weight, pI and extinction coefficient from a protein sequence, and concentration from A280.
- [All 39 lab calculators](https://www.labcanvas.com/tools/lab-calculators)

## Use this calculator from code or an AI agent

This calculator also answers as JSON, computed by the same code as this page: no key, no account, CORS open. Send the parameters as a query string (numbers may carry units; lists comma-separated) or POST them as JSON to https://www.labcanvas.com/api/tools/beer-lambert-law-calculator. An example, with the answer shortened to its first values:

```
GET https://www.labcanvas.com/api/tools/beer-lambert-law-calculator?absorbance=0.415&epsilon=6220
```

```json
{
 "ok": true,
 "tool": "beer-lambert-law-calculator",
 "inputs": {"absorbance":{"value":0.415},"epsilon":{"value":6220}},
 "result": {"concentration":{"label":"Answer","value":66.72025723,"unit":"µM"},"absorbance_value":{"label":"Absorbance","value":0.415}},
 "answer": "Concentration 66.72 µM (A = 0.415, 38.5 % transmitted).",
 "formula": "A = ε·l·c; A = −log₁₀ T = 2 − log₁₀ %T; molar and mass units joined by the molar mass (c[g/L] = c[mol/L] × M, ε[L g⁻¹ cm⁻¹] = ε[M⁻¹ cm⁻¹] / M)",
 "page": "https://www.labcanvas.com/tools/beer-lambert-law-calculator",
 "citation": "LabCanvas Beer–Lambert law calculator, https://www.labcanvas.com/tools/beer-lambert-law-calculator"
}
```

Parameters:

- `solve` (string, default concentration: concentration | absorbance | epsilon | path_length): The quantity to calculate (give the other three).
- `absorbance` (number): Absorbance A (or give transmittance).
- `transmittance` (number): Transmittance in % (A = 2 − log₁₀ %T), instead of absorbance.
- `epsilon` (number): Absorption coefficient, in epsilon_unit (e.g. 6220 for NADH at 340 nm).
- `epsilon_unit` (string, default M-1cm-1: M-1cm-1 | mM-1cm-1 | uM-1cm-1 | L/g/cm | mL/ug/cm | E1%): M-1cm-1 (L/mol/cm), mM-1cm-1, uM-1cm-1, L/g/cm (= mL/mg/cm), mL/ug/cm, or E1% (A of 10 mg/mL in 1 cm).
- `path_length` (number, cm, default 1): Path length (e.g. "1 cm", "1 mm").
- `concentration` (number, M): Concentration with its unit ("50 µM", "1 mg/mL"); a plain number is mol/L.
- `concentration_unit` (string, default µM: M | mM | µM | nM | g/L | mg/mL | µg/mL | ng/µL | mg/L | µg/L): Unit of the answer when solving for the concentration.
- `molar_mass` (number, g/mol): Molar mass (g/mol), needed when ε is molar and the concentration per mass, or the reverse.

All calculators: https://www.labcanvas.com/api/tools · OpenAPI 3.1: https://www.labcanvas.com/api/tools/openapi.json · MCP tool: `calc_beer_lambert_law` (LabCanvas MCP server).

## Questions

### How do I calculate concentration from absorbance?

Divide the absorbance by the molar absorption coefficient and the path length: c = A/(ε·l). With ε in M⁻¹ cm⁻¹ and l in cm, c is in mol/L. For example A = 0.415 and ε = 6220 M⁻¹ cm⁻¹ in a 1 cm cuvette give 6.67 × 10⁻⁵ M = 66.7 µM.

### What are the units of the molar absorption coefficient?

L mol⁻¹ cm⁻¹, written M⁻¹ cm⁻¹: absorbance is dimensionless, so ε = A/(l·c) has the units of 1/(cm × mol/L). 6.22 mM⁻¹ cm⁻¹ is 6220 M⁻¹ cm⁻¹. A mass (specific) coefficient is in L g⁻¹ cm⁻¹ (= mL mg⁻¹ cm⁻¹); divide a molar ε by the molar mass to get it.

### How do I convert transmittance to absorbance?

A = −log₁₀ T = 2 − log₁₀(%T). 100 %T is A = 0, 50 %T is A = 0.301, 10 %T is A = 1 and 1 %T is A = 2. Back: %T = 100 × 10^(−A).

### What is the extinction coefficient of NADH?

6220 M⁻¹ cm⁻¹ (6.22 mM⁻¹ cm⁻¹) at 340 nm (Horecker & Kornberg 1948); NAD⁺ does not absorb there, which is why NAD(P)H-linked enzyme assays are read at 340 nm.

### In what range is the Beer–Lambert law valid?

For dilute solutions of one absorbing species, measured with monochromatic light. Most readings are most precise for absorbances between about 0.1 and 1; above that stray light and detector limits make the response bend, and at high concentrations molecules interact. Dilute the sample or shorten the path, and check linearity with standards.

### How do I convert between M and mg/mL in the Beer–Lambert law?

Use the molar mass M (g/mol): c (mg/mL = g/L) = c (mol/L) × M, and ε per mass (mL mg⁻¹ cm⁻¹) = ε (M⁻¹ cm⁻¹) / M. Enter the molar mass in the calculator and it converts for you.

### What path length should I use?

The light path through the sample: 1 cm for a standard cuvette, the instrument’s value for microvolume readers (often 1 mm or 0.5 mm, sometimes normalised to 1 cm by the software), and for a microplate the liquid height, which depends on the volume in the well.

### What is E1%?

The absorbance of a 1 % (10 mg/mL) solution in a 1 cm path, often used for proteins. It is 10 × the mass coefficient in mL mg⁻¹ cm⁻¹, and ε (M⁻¹ cm⁻¹) = E1% × M / 10.