# Cyclic voltammetry analyser: peaks, ΔEp, E½ and Randles–Ševčík D

> Paste a CV or several scan rates: peak potentials and currents, ΔEp, E½, ipc/ipa and the Randles–Ševčík diffusion coefficient. Free, in your browser.

URL: https://www.labcanvas.com/tools/cyclic-voltammetry-analyzer

Paste potential and current (one or several scan rates, or an EC-Lab, Gamry, CH Instruments or Autolab export) and the analyser finds the anodic and cathodic peaks above a linear baseline, gives Epa, Epc, ipa, ipc, ΔEp = Epa − Epc, E½ = (Epa + Epc)/2 and the peak ratio, and over several scan rates fits ip against √v to get the diffusion coefficient from the Randles–Ševčík equation, ip = 0.4463 nFAC (nFvD/RT)^½, with the log–log slope that tells diffusion (0.5) from adsorption (1).

## Quick answer

- **What it does:** Finds the peaks of cyclic voltammograms (potentials, baseline-corrected currents, ΔEp, E½, peak ratio) and, over several scan rates, the diffusion coefficient by Randles–Ševčík and the log–log slope.
- **Formula:** ΔEp = Epa − Epc; E½ = (Epa + Epc)/2; ip = 0.4463 nFAC √(nFvD/RT), so D = (slope of ip vs √v)² RT / (0.4463² n³ F³ A² C²)
- **Example:** For the built-in simulated ferrocene-like couple (1 mM, 0.0707 cm², D = 2.4 × 10⁻⁵ cm²/s) at 20–500 mV/s: ΔEp = 56.3–60.5 mV, E½ = 0.430 V, ipa against √v has slope 92.66 µA/(V/s)^½ (R² = 0.9999), giving D = 2.38 × 10⁻⁵ cm²/s; log–log slope 0.505 (diffusion).

Free, in the browser, no account needed. Updated 2026-11-17.

## Features

- Paste potential and current (with a scan-rate column for several scan rates) or drop an EC-Lab .mpt, Gamry .DTA, CH Instruments or Autolab export; mV, µA, mA/cm² and mV/s are read from the column names
- Anodic and cathodic peak potentials from a parabola at the maximum, peak currents above a linear baseline extrapolated under the peak
- ΔEp, E½, |ipc/ipa| and Nicholson’s ipc/ipa, |Ep − Ep/2|, and a plain verdict: reversible, quasi-reversible or irreversible
- Randles–Ševčík plot (ip against √v) with fitted lines, D ± SE from the anodic and the cathodic peaks
- log|ip| against log v slope: 0.5 for diffusion, 1 for an adsorbed (surface-confined) species, with the surface coverage Γ
- Chooses the last complete cycle when the file has several, or the cycle you pick; scan rate from the time column when there is none
- Copy the peak table and a methods sentence; PNG of the voltammograms and of the Randles–Ševčík plot; CSV

## How to use it

1. **Paste or drop your voltammograms**: Potential (V), current and, for several scan rates, the scan rate (mV/s) on each row. A header row tells the units; without one, pick the current unit. Or drop the potentiostat’s export.
2. **Enter the electrode and the solution**: Electrons transferred n, electrode area (0.0707 cm² for a 3 mm disc) and the concentration in mM: they are needed for D, not for the peaks.
3. **Read the peaks**: Check the marked peaks on the plot, then read ΔEp, E½ and the peak ratio in the table.
4. **Read D**: With two or more scan rates, check that ip is straight against √v and the log–log slope is near 0.5, then read the diffusion coefficient.

## At a glance

| | |
| --- | --- |
| Input | Potential and current, with scan rate for several voltammograms; EC-Lab, Gamry, CH Instruments, Autolab exports, CSV |
| Peaks | Epa, Epc (parabola at the maximum), ipa, ipc above a linear baseline, ΔEp, E½, \|ipc/ipa\|, Nicholson ratio |
| Scan-rate study | ip vs √v (Randles–Ševčík D ± SE), log ip vs log v slope, surface coverage Γ for adsorbed species |
| Constants | F = 96485.33 C/mol, R = 8.31446 J/(mol·K); T editable (298.15 K) |
| Engine | The LabCanvas Graph app’s CV analysis |

## How it works

$$i_\mathrm{p} = 0.4463\, nFAC \left(\frac{nFvD}{RT}\right)^{1/2}$$

$$D = \frac{RT}{nF} \left(\frac{\mathrm{slope}}{0.4463\, nFAC}\right)^2$$

$$\Delta E_\mathrm{p} \approx \frac{2.218\,RT}{nF} = \frac{57}{n}\ \mathrm{mV\ at\ 25\,°C}$$

**Peaks.** Each sweep’s baseline is a straight line fitted to its start (2–15 % of the span) and extended under the peak; the peak current is measured from it, and the peak potential is the top of a parabola fitted within ±10 mV of the maximum. Anodic currents are positive (IUPAC). The reverse peak ratio is also given by Nicholson’s method, which corrects for the decaying forward current (Nicholson 1966).

**Reversible or not.** A reversible one-electron couple at 25 °C has ΔEp close to 57 mV (Bard & Faulkner, Table 6.2.1), independent of scan rate, and a peak ratio near 1. ΔEp that grows with scan rate means slow electron transfer (quasi-reversible); above about 200/n mV it is irreversible. Uncompensated resistance also widens ΔEp, so check it at low scan rates (Elgrishi et al. 2018).

**Diffusion coefficient.** For a freely diffusing species ip grows with √v, and the slope of ip against √v gives D from the Randles–Ševčík equation above (Bard & Faulkner eq. 6.2.19; 0.4463 nFAC(nF/RT)^½ = 2.69 × 10⁵ n 3/2 AC at 25 °C with A in cm², C in mol/cm³). The log–log slope separates diffusion (0.5) from an adsorbed layer (1), for which ip = n²F²vAΓ/4RT gives the surface coverage Γ (Laviron 1979) and D does not apply.

**Same numbers as the Graph app.** The page runs the LabCanvas Graph app's own CV analysis, which is checked against a Crank–Nicolson simulation, SciPy’s linregress and the Randles–Ševčík constant; the hand-off opens the same table and analysis there.

Sources: Bard AJ, Faulkner LR. *Electrochemical Methods*, 2nd ed. Wiley 2001. Nicholson RS. *Anal Chem* 1966;38:1406, [doi:10.1021/ac60242a030](https://doi.org/10.1021/ac60242a030). Laviron E. *J Electroanal Chem* 1979;101:19–28, [doi:10.1016/S0022-0728(79)80075-3](https://doi.org/10.1016/S0022-0728(79)80075-3). Elgrishi N et al. A practical beginner’s guide to cyclic voltammetry. *J Chem Educ* 2018;95:197, [doi:10.1021/acs.jchemed.7b00361](https://doi.org/10.1021/acs.jchemed.7b00361).

## Worked example: D of a ferrocene-like couple from five scan rates

The built-in example is a simulation, not a measurement: a reversible one-electron couple (E0′ = 0.43 V vs Ag/AgCl, D = 2.4 × 10⁻⁵ cm²/s, 1 mM) at a 3 mm disc (A = 0.0707 cm²), simulated by finite differences at 20, 50, 100, 200 and 500 mV/s with double-layer charging and noise.

| v (mV/s) | ipa (µA) | ipc (µA) | ΔEp (mV) | E½ (V) |
| --- | --- | --- | --- | --- |
| 20 | 12.86 | −12.31 | 60.5 | 0.4306 |
| 50 | 20.84 | −19.92 | 57.7 | 0.4318 |
| 100 | 29.51 | −28.36 | 57.4 | 0.4306 |
| 200 | 41.74 | −40.30 | 57.4 | 0.4298 |
| 500 | 65.37 | −63.52 | 56.3 | 0.4302 |

- **Reversible?** ΔEp stays at 56–61 mV at every scan rate, close to 57 mV, and E½ ≈ 0.430 V = E0′: a reversible couple.
- **Randles–Ševčík.** ipa against √v is a straight line with slope 92.66 ± 0.61 µA/(V/s) ½ (R² = 0.9999) through the origin.
- **D by hand.** With n = 1, F = 96485 C/mol, A = 0.0707 cm² and C = 1 mM = 10⁻⁶ mol/cm³: 0.4463 nFAC = 3.044 × 10⁻³; slope / 0.4463 nFAC = 92.66 × 10⁻⁶ / 3.044 × 10⁻³ = 0.03044; D = (RT/nF) × 0.03044² = 0.025693 × 9.264 × 10⁻⁴ = **2.38 × 10⁻⁵ cm²/s**. The simulation used 2.4 × 10⁻⁵.
- **Diffusion, not adsorption.** log|ipa| against log v has slope 0.505: the current is diffusion-controlled.



## Questions

### How do I calculate ΔEp and E½ from a cyclic voltammogram?

Read the anodic peak potential Epa and the cathodic peak potential Epc. ΔEp = Epa − Epc; E½ = (Epa + Epc)/2, which approximates the formal potential E0′ for a reversible couple. The analyser reads both from the peaks it finds.

### How do I calculate the diffusion coefficient from CV?

Record voltammograms at several scan rates, plot the peak current against √v and fit a line through the points. From the Randles–Ševčík equation, D = (RT/nF) × (slope / (0.4463 nFAC))², with A in cm² and C in mol/cm³. Check the line is straight and the log–log slope is about 0.5.

### What is the Randles–Ševčík equation?

For a reversible, diffusion-controlled couple at a planar electrode, ip = 0.4463 nFAC √(nFvD/RT); at 25 °C that is ip = 2.69 × 10⁵ n^(3/2) A C √(D v), with ip in A, A in cm², C in mol/cm³, D in cm²/s and v in V/s.

### How do I know if my redox couple is reversible?

ΔEp close to 57/n mV at 25 °C that does not change with scan rate, a peak ratio |ipc/ipa| near 1, and ip proportional to √v. ΔEp that grows with scan rate means quasi-reversible kinetics or uncompensated resistance.

### How are peak currents measured from the baseline?

Peak currents are measured from the extension of the current before the peak (the charging current and any earlier wave), not from zero. The analyser fits a straight baseline to the start of each sweep and extends it under the peak.

### What if my species is adsorbed on the electrode?

Then ip grows in proportion to v (log–log slope 1), ΔEp is small, and the Randles–Ševčík D does not apply. The analyser reports the surface coverage Γ from ip = n²F²vAΓ/4RT instead.