Photocatalytic degradation kinetics: rate constant k, half-life and % degradation
For photocatalytic degradation of a dye or pollutant, plot ln(C0/C) against irradiation time: for pseudo-first-order kinetics the points fall on a straight line whose slope is the apparent rate constant k_app, the half-life is ln 2/k_app, and the degradation is (1 − C/C0) × 100 %. C0 is the concentration when the light is switched on, after the dark adsorption–desorption step, not the concentration you started with. Paste time and concentration (or absorbance, proportional to concentration by Beer–Lambert) and the calculator finds C0 at light-on, fits the line, and gives k_app and t½ with 95 % confidence intervals, the removal in the dark and under light, and a second-order fit for comparison.
- What it does
- Fits the pseudo-first-order plot ln(C0/C) against time from light-on and gives the apparent rate constant, the half-life and the percentage degraded, keeping the dark adsorption step apart.
- Formula
- Example
- A 10 mg/L dye falls to 8.8 mg/L in 30 min in the dark (12.0 % adsorbed), then to 0.535 mg/L after 120 min of light: ln(C0/C) against t gives k_app = 0.02313 ± 0.00026 min⁻¹ (R² = 0.9991), t½ = 30.0 min (95 % CI 29.2–30.8 min) and 93.9 % photocatalytic degradation.
Loading Photocatalytic degradation kinetics calculator…
At a glance
| Model | Pseudo-first order: |
|---|---|
| Gives | k_app ± SE and 95 % CI, t½ = ln 2/k_app with 95 % CI, R², % degradation, removal in the dark |
| C₀ | The sample at light-on (after the dark adsorption step); time counted from light-on |
| Input | Time with concentration, absorbance (Beer–Lambert) or C/C₀; s, min or h |
| Comparison | Second order: |
| Output | C/C₀ and ln(C₀/C) plots (PNG), table and methods (copy), CSV; Graph app |
| Runs in | Your web browser; nothing is uploaded |
| Price | Free, no account needed |
What it does
How to use the photocatalytic degradation kinetics calculator
- Paste time and concentrationOne row per sample: the time and the concentration (or absorbance at the dye’s absorption maximum, or C/C₀), including the dark-phase samples and the one taken at light-on.
- Mark when the light went onEnter the time the lamp was switched on (0 when dark samples have negative times). Earlier samples are the dark adsorption step; C₀ is the sample at light-on.
- Read k and t½Read the apparent rate constant, the half-life and the percentage degraded, and check that ln(C₀/C) is a straight line.
- ReportCopy the methods sentence and the table, download the plots (PNG) and the CSV, or open the plot in the Graph app.
How it works
Pseudo-first order. The photocatalytic degradation of dyes and other organics on TiO₂ and similar catalysts is usually described by Langmuir–Hinshelwood kinetics, linregress) or through the origin. The 95 % confidence interval of
Dark adsorption and C₀. The suspension is usually stirred in the dark first, so the pollutant reaches adsorption–desorption equilibrium on the catalyst before irradiation (Houas et al. 2001). That drop is adsorption, not degradation: C₀ for the kinetics is the concentration at light-on, and time is counted from light-on. The calculator takes the sample at the light-on time as C₀ (or the last dark sample if there is none at that time) and reports the removal in the dark,
Absorbance instead of concentration. If you paste absorbance, the calculator uses
Second order, for comparison. Some data follow second-order kinetics,
Langmuir–Hinshelwood constants.
Worked example: a dye with 30 min of dark adsorption
The built-in example is synthetic, not a measurement: a 10 mg/L dye with catalyst, stirred in the dark for 30 min (to 8.8 mg/L), then irradiated for 120 min; the concentrations follow
| t (min) | C (mg/L) | Phase | ln(C₀/C) |
|---|---|---|---|
| −30 | 10.00 | dark | |
| −15 | 9.05 | dark | |
| 0 | 8.80 | light on: C₀ | 0 |
| 15 | 6.357 | light | 0.3252 |
| 30 | 4.348 | light | 0.7050 |
| 45 | 3.220 | light | 1.0054 |
| 60 | 2.147 | light | 1.4107 |
| 75 | 1.607 | light | 1.7004 |
| 90 | 1.088 | light | 2.0904 |
| 105 | 0.818 | light | 2.3756 |
| 120 | 0.535 | light | 2.8002 |
- Dark step:
mg/L at light-on, so was adsorbed before irradiation. - One point by hand: at 60 min,
. - Fit the line through the nine light points: slope
(95 % CI 0.02252–0.02375), intercept −0.0087, . The data were made with 0.023 min⁻¹. Forced through the origin, . - Half-life:
min (95 % CI 29.2–30.8 min). - Degradation:
after 120 min of light; with the dark step, of the dye was removed in all. Quoting 94.7 % as “photocatalytic degradation” would count the adsorption as reaction. - Second order, for comparison:
against curves upward (slope , ), so first order describes these data better.
If the sample at light-on is missing, the last dark sample (9.05 mg/L at −15 min) is taken as C₀ and the fit uses the eight light samples:
References
- Konstantinou IK, Albanis TA. TiO₂-assisted photocatalytic degradation of azo dyes in aqueous solution: kinetic and mechanistic investigations. Appl Catal B 2004;49:1–14. doi:10.1016/j.apcatb.2003.11.010
- Houas A, Lachheb H, Ksibi M et al. Photocatalytic degradation pathway of methylene blue in water. Appl Catal B 2001;31:145–157. doi:10.1016/S0926-3373(00)00276-9
- Kumar KV, Porkodi K, Rocha F. Langmuir–Hinshelwood kinetics: a theoretical study. Catal Commun 2008;9:82–84. doi:10.1016/j.catcom.2007.05.019
Frequently asked questions
How do I calculate the rate constant of photocatalytic degradation?
Plot ln(C0/C) against irradiation time, with C0 the concentration when the light was switched on, and fit a straight line: its slope is the apparent pseudo-first-order rate constant k_app (min⁻¹ if time is in minutes). This calculator does the fit and gives k_app with its standard error and 95 % confidence interval.
What is C0 in photocatalysis: the initial concentration or the one after the dark step?
For the kinetics, the concentration at light-on, after the adsorption–desorption equilibrium in the dark. The drop during the dark step is adsorption, not degradation. Report it separately as the removal in the dark.
How do I calculate the percentage degradation?
Degradation (%) = (1 − C/C0) × 100, with C0 at light-on and C at the end of irradiation. With absorbance, use (1 − A/A0) × 100, which is the same when absorbance is proportional to concentration.
Can I use absorbance instead of concentration?
Yes, if absorbance is proportional to concentration (Beer–Lambert): measure at the absorption maximum, within the linear range of a calibration, after removing the catalyst, and check that intermediates do not absorb there. ln(A0/A) then equals ln(C0/C), so k_app and the half-life are unchanged.
How do I calculate the half-life?
For first-order kinetics t½ = ln 2 / k_app = 0.693 / k_app. With k_app = 0.0231 min⁻¹, t½ = 30.0 min. For second-order kinetics t½ = 1/(k2·C0), which depends on C0.
Should the line go through the origin?
In theory ln(C0/C) = 0 at light-on, so a line through the origin is allowed; fitting an intercept as well shows whether the first minutes behave differently (an induction period, or fast initial adsorption). Report which you used; the two slopes are usually close.
What is the Langmuir–Hinshelwood model?
A rate law for reactions on a surface: r = k_r·K·C/(1 + K·C), with k_r the rate constant and K the adsorption constant. At low concentration it reduces to pseudo-first order with k_app = k_r·K. To get k_r and K, measure initial rates at several C0 and plot 1/r0 against 1/C0.
Why does my ln(C0/C) plot curve?
Common reasons: intermediates compete with the pollutant for the catalyst surface, the catalyst deactivates, the concentration is high enough that K·C is not small (Langmuir–Hinshelwood saturation), or absorbance is no longer proportional to concentration. Fit the early, straight part, and state the time range.
Link to this tool
Using it in a protocol, lab wiki or course page? Copy a link or citation.
<a href="https://www.labcanvas.com/tools/photocatalytic-degradation-kinetics">Photocatalytic degradation kinetics calculator</a> (LabCanvas)LabCanvas. Photocatalytic degradation kinetics calculator [online tool]. https://www.labcanvas.com/tools/photocatalytic-degradation-kineticsPlot the kinetics for your paper