LabCanvas
Free account · save your work, full figure editorSign up free

Mott–Schottky analyzer: flat-band potential and donor density

A Mott–Schottky plot is 1/C² against the applied potential, with C the space-charge capacitance per unit area. In the depletion region it is a straight line, 1/C² = 2/(e·ε·ε₀·N)·(E − E_fb − kT/e) for an n-type semiconductor, so the flat-band potential is the x-intercept minus kT/e (25.7 mV at 25 °C) and the donor density is N_D = 2/(e·ε·ε₀·slope); a negative slope means p-type, with E_fb = x-intercept + kT/e. Paste capacitance against potential (one or several frequencies) and the analyzer fits the straight part, gives E_fb and N with standard errors, and converts E_fb to the RHE scale.

What it does
Fits a straight line to 1/C² against potential over the linear (depletion) region and gives the flat-band potential and the donor or acceptor density for each frequency, with standard errors.
Formula
1C2=±2eεε0N(E−Efb∓kTe)Efb=E0∓kT/eN=2eεε0A2|slope|
Example
The built-in synthetic n-type film (ε = 86) at 1 kHz: the line from 0.077 to 0.977 V vs RHE has a slope of 8.525 × 10¹⁰ cm⁴ F⁻² V⁻¹ and meets the axis at 0.0748 V, so E_fb = 0.0748 − 0.0257 = 0.0491 V vs RHE and N_D = 1.92 × 10¹⁹ cm⁻³.

Last updated

FreeNo sign-upRuns in your browserData analysis

Loading Mott–Schottky analyzer…

At a glance

Mott–Schottky analyzer at a glance
Model1/C2=±2/(eεε0N)(E−Efb∓kT/e), C per cm²
GivesE_fb ± SE, N_D or N_A ± SE, type, slope, region, R²; per frequency
FitLeast squares over the linear region (automatic: longest run without significant curvature; or typed)
InputE with C (per cm² or whole electrode), 1/C², or Z″ with the frequency; optional frequency column
ConstantsCODATA 2018 (NIST): e, k, ε₀
OutputMott–Schottky plot (PNG 300/600 dpi), CSV, methods sentence, Graph app
Runs inYour web browser; nothing is uploaded
PriceFree, no account needed

What it does

How to use the mott–Schottky analyzer

  1. Paste or drop your dataColumns: potential and capacitance, with a frequency column if you measured several frequencies (rows one under another). A header such as E (V vs Ag/AgCl), C (µF/cm²), f (Hz) gives the units; without one, choose what the second column holds.
  2. Enter the semiconductor and the electrodePick the material for ε or type ε, and enter the electrode area if the capacitance is for the whole electrode.
  3. Check the frequency and the linear regionChoose the frequency to report and look at the plot: keep the straight part of 1/C², away from the bend near E_fb and from the high-bias end.
  4. Choose the potential scaleKeep the measured scale, or pick the reference electrode and the pH to report E_fb against RHE.
  5. Read and exportRead E_fb, N and the type, compare the frequencies, and copy the results, the methods sentence or the plot.

How it works

1Csc2=2eεε0A2ND(E−Efb−kTe)

The Mott–Schottky equation. At a semiconductor–electrolyte junction under depletion, the space-charge layer behaves as a capacitor whose thickness grows with the band bending, so 1/Csc2 rises linearly with the potential. Gelderman, Lee and Donne (2007) write it as above for an n-type electrode, with Csc the space-charge capacitance, A the electrode area, ε the relative permittivity of the semiconductor, ε0 the vacuum permittivity, ND the donor density, e the elementary charge, k Boltzmann’s constant and T the temperature; the same relation is derived in Bard and Faulkner’s textbook (section 18.2). For a p-type electrode the slope is negative and the sign of kT/e changes: 1/C2=−2/(eεε0A2NA) ×(E−Efb+kT/e).

What the analyzer computes. It converts every row to the capacitance per cm² (dividing by the area when the capacitance is for the whole electrode; from impedance as C=1/(2πf|Z″|), a series RC), takes y=1/C2 in cm⁴ F⁻², and fits y=a+bE by least squares over the linear region. The x-intercept is E0=−a/b; then

Efb=E0∓kTe,N=2eεε0|b|

(minus for n-type, plus for p-type, the type from the sign of b), with ε0 in F/cm so that N is in cm⁻³. With the capacitance per cm² the area is already in C, which is the same as the A2 in the formula above. Standard errors: of N from the standard error of the slope, and of Efb by the delta method from the variances and covariance of a and b. Constants are the CODATA 2018 values (NIST): e=1.602176634×10−19 C and k=1.380649×10−23 J/K (exact), ε0=8.8541878128×10−12 F/m; kT/e=25.693 mV at 25 °C.

The linear region. Near Efb the depletion approximation fails and the plot bends; at high bias, surface states, a Helmholtz layer in series or the thickness of a thin film bend it too. Automatically, the analyzer takes the longest run of consecutive points (at least five, and a quarter of the data) in which a quadratic term does not improve the straight line significantly (F test, P ≥ 0.01), the rule of the LabCanvas Graph app’s Mott–Schottky analysis, so both give the same numbers. You can type the range instead.

Frequency. The capacitance of a real electrode is usually measured at several frequencies (for example 0.5, 1 and 2 kHz). If the space-charge capacitance is what you measure, the lines meet the axis at the same potential, though the slopes may differ; when E_fb moves with the frequency, the result is uncertain and the frequency should be reported. The analyzer fits every frequency and lists them side by side.

RHE scale. E(RHE) = E + E°(ref) + (ln 10·RT/F)·pH, with the reference electrodes and values of the RHE converter. The slope of 1/C² does not change; only the potentials shift.

Worked example: an n-type film measured at three frequencies

The built-in example is synthetic, not a measurement: an n-type film with ε = 86 (the rutile value of Parker 1961, perpendicular to the c axis), N_D = 2 × 10¹⁹ cm⁻³ and E_fb = −0.950 V vs Ag/AgCl (saturated KCl), measured at 0.5, 1 and 2 kHz in an electrolyte of pH 14 at 25 °C, from −1.100 to −0.050 V in 25 mV steps. To make it realistic, the curve is rounded near E_fb, a 100 µF/cm² Helmholtz layer is in series, the capacitance falls slightly with frequency (C ∝ f^−0.03), and there is 0.4 % noise. The capacitance is per cm², so no area is needed.

  1. Convert to RHE. E(RHE)=E+0.199 +0.05916×14 =E+1.0272 V.
  2. Take 1/C². At −0.500 V vs Ag/AgCl and 1 kHz, C = 5.0511 µF/cm², so 1/C2=1/(5.0511×10−6)2 =3.919×1010 cm4F−2.
  3. Fit the straight part. At 1 kHz the automatic region runs from 0.077 to 0.977 V vs RHE (37 points): slope b=(8.525±0.036)×1010 cm⁴ F⁻² V⁻¹, intercept a=−6.376×109 cm⁴ F⁻², R² = 0.99936.
  4. Flat-band potential. E0=−a/b=0.0748 V; the slope is positive, so n-type, and Efb=0.0748−0.0257 =0.0491±0.0022 V vs RHE (−0.978 V vs Ag/AgCl).
  5. Donor density. ND=2/(eεε0b) =2/(1.6022×10−19 ×86 ×8.8542×10−14 F/cm ×8.525×1010) =1.92×1019 cm−3.
The example at each frequency (automatic region 0.077–0.977 V vs RHE)
FrequencyE_fb (V vs RHE)N_D (cm⁻³)R²
500 Hz0.04941.99 × 10¹⁹0.99923
1 kHz0.04911.92 × 10¹⁹0.99936
2 kHz0.05061.84 × 10¹⁹0.99931

The three lines meet the axis within 1.5 mV of each other while the slopes differ by 8 %: the dispersion changes N_D, not E_fb. The fitted E_fb is 28 mV below the true 0.077 V vs RHE, because the Helmholtz layer in series and the rounding near E_fb both bend 1/C². That is typical: a Mott–Schottky flat-band potential carries an uncertainty of some tens of millivolts beyond its standard error, so state the frequency, the region fitted and ε with the result.

Relative permittivity ε of common semiconductors

N is inversely proportional to ε, so the value you use matters as much as the slope. These are static values of single crystals near room temperature, each from the paper cited; films, nanostructures and porous or doped oxides can differ a lot, and for many oxides used as photoelectrodes the published values spread widely. Use a value measured for your material if you can, and cite it.

Static relative permittivity near 25 °C
MaterialεNoteSource
Si11.711.7 ± 0.2 at 1 MHzDunlap & Watters 1953
GaAs13.0912.35 × (1 + 2.01 × 10⁻⁴ K⁻¹ × 298.15 K)Strzalkowski et al. 1976
CdTe11.0110.31 × (1 + 2.27 × 10⁻⁴ K⁻¹ × 298.15 K)Strzalkowski et al. 1976
ZnSe9.258.80 × (1 + 1.71 × 10⁻⁴ K⁻¹ × 298.15 K)Strzalkowski et al. 1976
TiO₂, rutile ⊥ c86a direction, 300 KParker 1961
TiO₂, rutile ∥ c170c direction, 300 KParker 1961

Strzalkowski et al. give ε extrapolated to 0 K and a linear temperature coefficient λ (valid from 100 to 300 K); the values above are ε(0)(1+λT) at 298.15 K.

Common Mott–Schottky mistakes

  • Capacitance of the whole electrode used as if per cm². N scales with 1/A²: a 0.2 cm² electrode read as 1 cm² gives N 25 times too high. Enter the area, or the capacitance per cm².
  • Forgetting kT/e. The x-intercept is E_fb + kT/e for n-type (E_fb − kT/e for p-type); at room temperature the difference is 25.7 mV.
  • Fitting the bend. Points near E_fb and at high bias are not on the depletion line; including them moves the intercept.
  • One frequency, no check. If E_fb changes with the frequency, the capacitance is not only the space charge’s (surface states, roughness, a porous film); report the frequency or the spread.
  • Mixing potential scales. Convert to RHE (or state the reference) before comparing E_fb with band positions or other papers.

References

  1. Gelderman K, Lee L, Donne SW. Flat-band potential of a semiconductor: using the Mott–Schottky equation. J Chem Educ 2007;84:685. doi:10.1021/ed084p685
  2. Bard AJ, Faulkner LR. Electrochemical Methods: Fundamentals and Applications, 2nd ed. Wiley; 2001, section 18.2 (semiconductor electrodes). ISBN 978-0-471-04372-0.
  3. Tiesinga E, Mohr PJ, Newell DB, Taylor BN. CODATA recommended values of the fundamental physical constants: 2018. Rev Mod Phys 2021;93:025010. doi:10.1103/RevModPhys.93.025010 (NIST: physics.nist.gov/cuu/Constants)
  4. Dunlap WC, Watters RL. Direct measurement of the dielectric constants of silicon and germanium. Phys Rev 1953;92:1396–1397. doi:10.1103/PhysRev.92.1396
  5. Strzalkowski I, Joshi S, Crowell CR. Dielectric constant and its temperature dependence for GaAs, CdTe, and ZnSe. Appl Phys Lett 1976;28:350–352. doi:10.1063/1.88755
  6. Parker RA. Static dielectric constant of rutile (TiO₂), 1.6–1060 °K. Phys Rev 1961;124:1719–1722. doi:10.1103/PhysRev.124.1719
  • RHE converterPotentials vs Ag/AgCl, SCE or Hg/HgO converted to the RHE scale, with iR correction and overpotential.
  • Tauc plot band gap calculatorOptical band gap from a UV–vis or diffuse-reflectance spectrum by a Tauc plot, with its 95% CI.
  • Tafel slope calculatorTafel slope in mV/dec and the exchange current density from a polarisation curve.
  • All 39 lab calculatorsSolutions, molecular biology, cell culture, assays, electrochemistry and materials, on one page.

Frequently asked questions

How do you calculate the flat-band potential from a Mott–Schottky plot?

Plot 1/C² (C per unit area) against the potential, fit a straight line to the linear part and find where it meets the potential axis, E₀ = −intercept/slope. For an n-type semiconductor (positive slope) the flat-band potential is E₀ − kT/e; for p-type (negative slope) it is E₀ + kT/e. kT/e is 25.7 mV at 25 °C.

How do I calculate the donor density from the Mott–Schottky slope?

N_D = 2/(e·ε·ε₀·slope), with the slope of 1/C² against E in cm⁴ F⁻² V⁻¹, ε the relative permittivity, ε₀ = 8.854 × 10⁻¹⁴ F/cm and e = 1.602 × 10⁻¹⁹ C; N_D comes out in cm⁻³. If C is the capacitance of the whole electrode, divide also by A² (A in cm²). A slope of 8.5 × 10¹⁰ cm⁴ F⁻² V⁻¹ with ε = 86 gives 1.9 × 10¹⁹ cm⁻³.

Does a positive or negative slope mean n-type or p-type?

A positive slope of 1/C² against potential means n-type (donors): anodic bias deepens the depletion layer. A negative slope means p-type (acceptors). The analyzer reads the type from the sign and reports N_D or N_A.

Which frequency should I use for Mott–Schottky?

Measure at several frequencies in the range where the capacitance is that of the space charge, often around 0.5–10 kHz, and check that the flat-band potential does not depend on the frequency. Paste them all with a frequency column: the analyzer fits each one and compares them.

What dielectric constant should I use?

The static relative permittivity of your semiconductor, ideally measured for the same kind of material. The table on this page gives referenced single-crystal values (Si 11.7, GaAs 13.1, rutile TiO₂ 86 or 170); thin films and nanostructures can differ, so state the value and its source with your result.

Why is my Mott–Schottky plot not linear?

Near the flat-band potential the depletion approximation fails, and at high bias surface states, a Helmholtz layer in series, deep depletion or a thin film fully depleted bend the plot. A rough or porous electrode also gives frequency-dependent capacitance. Fit only the straight part and report the range.

How do I convert the flat-band potential to the RHE scale?

E(RHE) = E(ref) + E°(ref) + 0.0592 V × pH at 25 °C: for Ag/AgCl in saturated KCl (0.199 V) at pH 14, add 1.027 V. Choose the reference and the pH in step 4, or use the RHE converter.

Can I use the capacitance from an impedance fit?

Yes: paste the space-charge capacitance from your equivalent-circuit fit against potential. If you only have the imaginary impedance at one frequency, paste Z″ with the frequency and the analyzer takes C = 1/(2πf|Z″|), which assumes a simple series RC.

Plot the Mott–Schottky data for your paper

The LabCanvas Graph app draws the Mott–Schottky plot with the fitted line and the flat-band potential, from the same analysis, ready to restyle and size for a journal. Your data from this page open in it.

Open the Mott–Schottky graph